High-strength engineering plastic rack, light, beautiful and corrosion resistant
Adult, child
O2 0.27--0.55MPa
O2: 0.1--1.0L/min; 1.1--10L/min
There will be sound alarm when the oxygen pressure < 0.2MPa
It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0-5vol%.
Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer.
Working mode: all-close, semi-close, semi-open
Releasing pressure: 0.5-7kPa.
Bellows for adults, bellows for children, tidal volume range:0-1500ml
High-definition 5.7” LCD screen display
IPPV, MANUAL (Manual monitoring of tidal, volume, ventilation volume, respiratory rate).
Tidal volume 50--1500ml
Rate 2--99bpm
SIMV rate 2--20bpm
I:E 2:1--1:8
Inspiratory trigger pressure -1.0--2.0kPa
Pressure range 1.0--6.0kPa
Tidal volume, ventilation volume, IPPV rate, I/E, peak pressure of airway, inspiratory trigger pressure, pressure - time waveform.
Upper limit setting range 1.0--6.0kPa;
Low limit setting range 0.4--2.0kPa.
Per-minute Ventilation Volume Alarm
Upper limit setting range 3.0--30L/min;
Low limit setting range 1.0--10L/min.
It will give alarm when stress have consistently been higher than 2.5 kPa
It will give sound and light alarm if there is no tidal volume input in 15s
AC 220V±10% 50Hz±1Hz