High-strength engineering plastic rack, light, beautiful and corrosion resistant
Adult, child
O2: 0.27--0.55Mpa;
N2O: 0.27--0.55MPa.
O2: 0.1--10L/min;
N2O: 0.1--10L/min
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure < 0.2MPa, the flow of nitrous oxide would be cut off.
25--75L/min
There will be sound alarm when the oxygen pressure < 0.2MPa
It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0-5vol%.
Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.
Working mode: all-close, semi-close, semi-open;
Releasing pressure: 0.5-7kPa.
Respiratory Bellows
Bellows for adults, bellows for children, tidal volume range 0-1500ml
High-definition 5.7” LCD screen display
IPPV, SIPPV, SIMV, PEE, MANUAL, SIGH
Tidal volume 50--1500ml
Rate 2--99bpm
SIMV rate 2--20bpm
I:E 2:1--1:8
Inspiratory trigger pressure -1.0--2.0kPa
PEEP 0--2.0kPa
Pressure range 1.0--6.0kPa
SIGH 1.5 times the inspiratory time
(21%--100%)
Oxygen Concentration alarm
Upper limit setting range (50%--100%)
Low limit setting range (15%--50%)
Upper limit setting range 1.0--6.0kPa;
Low limit setting range 0.4--2.0kPa.
Per-minute ventilation volume alarm
Upper limit setting range 3.0--30L/min;
Low limit setting range 1.0--10L/min.
It will give alarm when stress have consistently been higher than 2.5 kPa
It will give sound and light alarm if there is no tidal volume input for 15 seconds
AC 220V±10%, 50Hz±1Hz, (Storage battery, UPS)
Note: Items within brackets are optional.